ProductDescription
2 MarathonSeriesFA/FR
2ProductDescription
TheMarathonFA/FRfiberopticseriesofinstrumentsarehigh‐performanceinfraredthermometers.
Eachhasafrontendconsistingofasmall,fixedfocusopticalheadcoupledtoaruggedfiberoptic
cablewrapped withaflexiblestainlesssteelsheath.Thefiber opticcableattachesto anelectronics
enclosure,whichcan
bemountedawayfromthehot,hostileenvironment.Theelectronicsenclosure
canbeconnectedtoacomputerwithitstwo‐wayRS485interface.
Temperaturemeasurementscanbetakenusingeitherofthefollowingmodes:
• 1‐color mode (FA and FR sensors) – for standard temperature measurements. The 1‐color
mode
isbestformeasuringthetemperatureoftargetsinareaswherenosightingobstructions,
eithersolidorgaseous,exist.The1‐colormodeisalsobestwherethetargetcompletelyfills
themeasurementspotandwherethebackgroundorforegroundarehigherintemperature
thanthetarget.
• 2‐colormode
(FRsensorsonly)–temperaturesaredeterminedfromtheratiooftwoseparate
andoverlappinginfraredbands.The2‐colormodeisbestformeasuringthetemperatureof
targets that are partially obscured (either intermittently or permanently) by other objects,
openings,screens,orviewingwindowsthatreduceenergy,andby
dirt,smoke,orsteamin
theatmosphere.The2‐colormodeca nalsobeusedontargetsthatdonotcompletelyfillthe
measurementspot,providedthebackgroundismuchcoolerthanthetarget.
Eachmodeloperatesasatemperaturemeasurementsubsystemconsistingofopticalelements,spectral
filters,detector,and
digitalelectronics.Allcomponentsarewater‐tightNEMA‐4(IEC529,IP65)rated
andarebuilttooperateona100percentdutycycleinindustrialenvironments.Simultaneousanalog
and digital outputs consist of standardized current signals commonly available for use with
computers,controllers,recorders,alarms,orA/Dinterfaces.
Model Description
FA1A, FA1B, FA1C 1-color-sensor in spectral range of 1 µm
different temperature ranges
FA1G 1-color-sensor specifically designed for measuring glass
FA2A, FA2B 1-color-sensor in spectral range of 1.6 µm
different temperature ranges
FR1A, FR1B, FR1C 2-color-sensor in spectral range of 1 µm (nominal)
different temperature ranges
Table1:Models
2.1TheoryofOperationfor2‐ColorSensors
Two‐colorratiotechnologymakespossibleaccurateandrepeatabletemperaturemeasurementsthat
are freefrom dependence on absolute radiated energyvalues. In use, a 2‐color sensor determines
temperaturefromtheratiooftheradiatedenergiesintwoseparatewavelengthbands(colors).The
benefits of 2‐color sensors are that accurate
measurements can be made under the following
conditions:
• Whenthefieldofviewtothetargetispartiallyblockedorobscured.
• Whenthetargetissmallerthanthesensor’sfieldofview.
• Whentargetemissivitiesareloworchangingbythesamefactorinbothwavelengthbands.