RayTek FR Thermometer User Manual


 
Maintenance
MarathonSeriesFA/FR 49
10.4ReplacingtheFiberOpticCable
FAfibercableassembliesarenotfieldʺreplaceableʺwithoutblackbodyrecalibration!
Assuch,spareFAfibercableassembliesarenotavailable!
Ifthefiberopticcableeverneedstoberemovedorreplaced,itcanberemovedfromboththeoptical
headandelectronicsenclosurewithoutdemountingthemfromtheir
brackets.
Pleasebeawareofthefollowingwhenremovingorinstallingcables:
Makesurecableconnectorsatthesensingheadandelectronicsenclosurearecleanbefore
removingand/orreplacingthefiberopticcable.
Replacementfiberopticcablesofthesamelengthcanberecalibratedinthefieldbyusingthe
suppliedFiberReplacementCalibrationsoftware.Replacementfiberopticcablesofdifferent
lengthsrequirerecalibrationatthefactory,oratafactoryauthorizedservicecenter.Contact
yoursalesrepresentativefordetails.
Always clean the area around the fiber optic cable connectors before disconnecting. If any
contaminants get into the open connectors,
the sensor’s accuracy will be compromised. After
removingthecable,orbeforeinstallinganewcable,theendsmustbeprotectedatalltimesuntil
connectedtothesensingheadandelectronicsenclosure.Cablesareshippedwithprotectiveendcaps.
Always save these caps for use whenever the fiber optic cable
must be disconnected. Any
contaminationtothefiberopticcableendswilldegradeperformance.Toreplacethefiberopticcable,
youwillneedtodisconnectitfromboththeopticalheadandtheelectronicsenclosure.Thefollowing
instructionswillguideyouthroughtheprocess.
10.4.1RemovingtheFiberOpticCable
10.4.1.1RemovingtheFiberOpticCablefromtheOpticalHead
Completethefollowingstepstodisconnectthefiberopticcablefromtheopticalhead:
1. Thoroughlycleantheareaaroundtheopticalhead.
2. Inserta1.3mm(0.050”)hexwrenchintotheopticalhead’shexscrewandturncounter
clockwise
untilthecableisloose.
3. Drawthefiberopticcableoutoftheopticalhead.
4. ImportantIfyouplantoreconnectthesamecable,immediatelycovertheendwithaslipon
endcaptopreventcontamination.Donotuseanyadhesivetapeoverthecableend.