RayTek FR Thermometer User Manual


 
Operation
MarathonSeriesFA/FR 27
Figure25:ModeSelectorButtonSequence(FRModels)
6.2.1TemperatureDisplay
Thetemperaturedisplaycanbesetforeither°Cor°FbypressingtheC/Fselectorbutton(cup
arrow),whichalsodoublesastheIncreaseValuebuttonfortheothermodes.TheDecreaseValue
(d downarrow)buttonisinactiveinthismode.AlitLEDshows
youwhetherthemeasured
temperatureisin°Cor°F.NotethatthissettinginfluencestheRS485outputforbothtargetand
internaltemperatures.
6.2.2Emissivity(1Color)
Youcansettheunitupforeither1coloror2colormeasurements.The1C/2Cselectorbuttononthe
controlpanelswitchesbetweenthetwofunctions.OneoftheredLEDs,labeled1Cand2C,willshow
whatfunctionisactive.
Theemissivityisacalculatedratioofinfraredenergy
emittedbyanobjecttotheenergyemittedbya
blackbodyatthesametemperature(aperfectradiatorhasanemissivityof1.00).Theemissivityis
presetat1.00.Forinformationondetermininganunknownemissivity,andforsampleemissivities,
refertosection
11.1DeterminationofEmissivityand11.2TypicalEmissivityValues,p.54f.
Tochangetheunit’semissivitysetting,completethefollowing:
1. Makesurethe1CLEDislit.
2. PresstheModebuttonuntiltheЄLEDislit.Thecurrentemissivityvalueshowsonthe
display.
3. Pressthec ord buttontochangethevalue.
4. Pressthe
ModebuttonseveraltimesuntiltheCorFLEDislit.Thedisplayedtemperature
willnowbebasedonthenewemissivityvalue.
6.2.3Slope(2Color)
Theslopeisthequotientoftheemissivitiesbasedonthenarrowandthewidespectralrange(firstand
secondwavelength).Theslopeispresetatthefactoryat1.000.
Display/Change emissivity (1-color mode)
default: 1.00
Display/Change slope (2-color mode)
default: 1.000
Display current temperature Switches between °C and °F
Raises and lowers emissivity
or slope
Display/Change peak hold setting
default = 0 sec / off
Raises and lowers
peak hold timing
Display/Change averaging setting
default = 0 sec / off
Raises and lowers
averaging time