RayTek FA Thermometer User Manual


 
Maintenance
MarathonSeriesFA/FR 47
Therelayiscontrolledbythetemperatureselectedonthedisplay.Ifanyfailsafecodeappearsonthe
display,therelaychangestothe“abnormal”state.Theexceptionisthe“dirtywindow”condition.
Thiscausestherelaytochangestate,leavinganormalnumericaltemperatureoutput.Thedirty
windowisdetected
ineither1coloror2colormode.
Error Code
0 – 20 mA Output 4 – 20 mA Output
no error according to temperature
according to temperature
ECHH 21 to 24 mA 21 to 24 mA
ECUU 0 mA 2 to 3 mA
EIHH 21 to 24 mA 21 to 24 mA
EIUU 0 mA 2 to 3 mA
EUUU 0 mA 2 to 3 mA
EHHH 21 to 24 mA 21 to 24 mA
EAAA 0 mA 2 to 3 mA
Table11:CurrentOutputValuesinaccordancetoanError
Iftwoerrorsoccursimultaneously,thehigherpriorityerroristheonethatispresentedontheLED’s
digitalandanalogoutputs.Forexample,in2colormode,iftheinternalambientistoohighandthe
attenuationistoohigh,
theunitoutputsEIHHontheLED’sanddigitaloutputand21mAonthe
analog output. However, since 2color wide band and narrow band temperatures may all be
presentedsimultaneouslythroughRS485,theiroverandunderrangeconditionsareindependent.
Examplesoffailsafeconditions:
1. Onecolor temperature
is selected for display on the LED’s. Twocolor temperature is
transmittedinburstmode.Widebandtemperatureisunder range.Twocolortemperatureis
999°C.
Outputs:
Display: EUUU
RS485: CT0999
Analog: 2to3mA
Relay: abnormalstate
2. Twocolor temperature is selected for display
on LED’s. All three temperatures are
transmitted in burst mode. Twocolor temperature is 1021°C. Wide band temperature is
703°C.Narrowbandtemperatureis685°C.Attenuationisabove95%,the“dirtywindow”
threshold.
Outputs:
Display: 1021
RS485: CT1021W0703N0685
Analog: scaledtotemperature,between4and
20mA
Relay: abnormalstate