RayTek FA Thermometer User Manual


 
Operation
28 MarathonSeriesFA/FR
Theslopeisthedecidingparameterformeasurementsin2colormode!Theemissivity
affectsonlymeasurementsin1colormode.
Forinformationondetermininganunknownslope,andforsampleslopes,refertosection
11.3Typical
Slopes,p.
56.
Tochangetheunit’sslopesetting,completethefollowing:
1. Makesurethe2CLEDislit.
2. PresstheModebuttonuntiltheЄLEDislit.Thecurrentslopevalueshowsonthedisplay.
3. Pressthec ord buttontochangethevalue.
4. Pressthe
ModebuttonseveraltimesuntiltheCorFLEDislit.Thedisplayedtemperature
willnowbebasedonthenewslopevalue.
6.2.42C/1CSwitch
Toswitchbetween2colorand1colortemperaturemeasurementpushthe2C/1Cselectorbutton.Alit
LEDindicatestheactivemeasurementmethod.SwitchingaffectstheLEDdisplayandanalogoutbut
nottheRS485out.
6.2.5PeakHold(PKH)
WithPeakHold,therespectivelastpeakvalueisheldforthedurationofHoldTime.
TosetandactivatePeakHold,dothefollowing:
1. PresstheModebuttonuntilthePKHLEDislit.
2. Pressthecbuttontobothsetandactivate.Thedisplayreadsin0.1
seconds.SetPeakHold
from0.1to299.9seconds.IfPeakHoldissetto300.0seconds,ahardwareresetisneededto
triggeranotherreading.IfPeakHoldissetto0.0seconds,thefunctionisdeactivated.
3. PresstheModebuttonuntiltheCor FLEDislit.
IfPeakHoldhasbeenactivated,thePeak
LEDwillstaylit.
OncePeakHoldissetabove0,itautomaticallyactivates.Theoutputsignalremainsthesameuntilone
oftwothingshappens:
Thepeakholdtimerunsout.Inthiscase,thesignalrevertstoactualtemperature.
Theactualtemperaturegoesabovetheholdtemperature.Inthiscase,startsholdingnew
peak.
Notethatotherholdfunctions(likeValleyHoldorAveraging)cannotbeusedconcurrently.
Bymeansofthesoftwareotherholdfunctionsareadjustable(e.g.AdvancedPeakHold).
6.2.6Averaging(AVG)
Averagingcanbeusefulwhenanaveragetemperatureoveraspecificdurationisdesired,orwhena
smoothingoffluctuatingtemperaturesisrequired.
The averagingalgorithm simulatesa first order lowpass RCfilter whosetime constantcan be
adjustedtomatchtheuser’saveragingneeds.Thefollowingfigureillustratesan
averagingoutput
signal.