SensorLocation
MarathonSeriesFA/FR 13
4SensorLocation
Sensorlocationandconfigurationdependsontheapplication.Beforedecidingonalocation,youneed
tobe awareofthe ambienttemperature ofthe location,theatmosphericqualityofthe location
(especiallyfor1colortemperaturemeasurements),andthepossibleelectromagneticinterferencein
thatlocation(aconsiderationonlyfortheelectronics
enclosure).If youplantouseairpurging,you
needtohaveanairconnectionavailable.Also,wiringandconduitrunsmustbeconsidered,including
computerwiringandconnections,ifused.Thefollowingsubsectionscovertopicstoconsiderbefore
youinstallthesensor.
4.1AmbientTemperature
Theopticalheadisdesignedtooperateinambienttemperaturesupto200°C(390°F).Theelectronics
enclosureis designedtooperateinambienttemperaturesbetween0°C(32°F)and60°C(140°F).The
internalambienttemperaturecanvaryfrom10°C(50°F)to68°C(154°F).Internaltemperaturesoutside
thisrangewillcauseafailsafe
error.
4.2AtmosphericQuality
Smoke,fumes,dust,andothercontaminantsintheair,aswellasadirtylensaregenerallynota
problemwhenusingthe2colormode(aslongastheattenuationisequalinbothspectralbands).
However,ifthelensgetstoodirty,itcannotdetectenoughinfraredenergyto
measureaccurately,and
theinstrumentwillindicateafailure.Itisgoodpracticetoalwayskeepthelensclean.TheAirPurge
Collarhelpskeepcontaminantsfrombuildinguponthelens.Ifyouuseairpurging,makesureanair
supplywiththecorrectairpressureisinstalledbefore
proceedingwiththesensorinstallation.
4.3ElectricalInterference
Tominimizeelectricalorelectromagneticinterferenceor“noise”beawareofthefollowing:
Mounttheelectronicsenclosureasfarawayaspossiblefrompotentialsourcesofelectrical
interferencesuchasmotorizedequipmentproducinglargesteploadchanges.
Useshieldedwireforallinputandoutputconnections.
Makesurethe
shieldwirefromtheelectronicstoterminalblockcableisearthgrounded.
Foradditionalprotection,useconduitfortheexternalconnections.Solidconduitisbetterthan
flexibleconduitinhighnoiseenvironments.
DonotrunACpowerforotherequipmentinthesameconduit.
Wheninstalling the optical head, check forany highintensity discharge lampsor
heatersthatmaybeinthefieldofview(eitherbackgroundorreflectedonashiny
target)!Reflectedheatsourcescancauseasensortogiveerroneousreadings.
4.4DistancetoObject
The requested spot size determines the maximum distance to the measurment object and the
necessaryfocusoftheoptic.TheStandardFocusissetatinfinity.TheCloseFocusopticalheadsare
focusedat100mm(4in)or300mm(12in),seesection
3.5OpticalSpecifications,p.9.