RayTek FA Thermometer User Manual


 
ProgrammingGuide
MarathonSeriesFA/FR 41
SensorModelType
SensorSerialNumber
RelayControl
Laserstatus
Setpoints
Deadband
CurrentOutputMode(0‐20mAor4‐20mA)
Thefollowingitemscannotbepolled:
Poll/BurstMode
Baudrate
Relaycontrol
Setcurrentoutput
Anexamplestringforcommand$=UTQEGH<CR>
Thedefault
stringisasfollows:CT1250Q0400.023E1.00G005.5H1400<CR><LF>
9.4ResponseTimeinSetupMode
Theanalogoutputresponsetimeisnotguaranteedwhileaparametervalueisbeingchangedorif
thereisacontinuousstreamofcommandsfromthehost.
ThedigitalresponsetimespecifieshowquicklytheunitcanreportatemperaturechangeviaRS485in
burstmode.(Digitalresponsetimeisnot
definedforpolledmode.)Thedigitalresponsetimeis
definedasthetimethatelapsesbetweenachangeintargettemperatureandthetransmissionofa
burststringreportingthenewtemperature.Actualdigitalresponsetimecanvaryfromonereadingto
thenext,sothedigitalresponsetimeis
definedasthe“averagedigitalresponsetime.”
Theaveragedigitalresponsetimedependsonthenumberofcharactersrequestedintheoutputstring
andwiththebaudrate.Itmaybecomputedasthefollowing:
b
n
t
15000
9.9
+=
where:
t=averageresponsetimeinms
n=thenumberofcharactersinthestringincluding<CR>and<LF>
b=thebaudrate
Example:
Withabaudrateof38400,andanoutputstringcontainingtemperatureunits,2colortemperatureand
ambient(20characters),theaveragedigital
responsetimewouldbethefollowing:
mst 7.17
38400
1500020
9.9 =
+=
Notethattheanalogoutputresponseti meisnotaffectedbybaudrateorthenumberofcharacters
transmittedintheburststring.