RayTek FA Thermometer User Manual


 
Operation
MarathonSeriesFA/FR 29
Figure26:AveragingExample
TosetandactivateAveraging,dothefollowing:
1. PresstheModebuttonuntiltheAVGLEDislit.
2. Pressthecbuttontobothsetandactivate.Thedisplayreadsin0.1seconds.SetAverage
anywhere from 0.1 to 300.0 seconds. If Average is set
to 0.0 seconds, the function is
deactivated.
3. PresstheModebuttonuntiltheCorFLEDislit.IfAveragehasbeenactivated,theAVGLED
willstaylit.
OnceAveragingissetabove0,itautomaticallyactivates.Notethatotherholdfunctions(likePeak
HoldorValley
Hold)cannotbeusedconcurrently.
6.2.7ValleyHold(VAL)
WithValleyHold,therespectivelastvalleyvalueisheldforthedurationofHoldTime.
FunctionandsettingforvalleyholdcorrespondstothealreadydescribedPeakHoldfunction,see
section
6.2.5PeakHold(PKH),p.28.TheValleyHoldfunctionisnotavailablefor2colorunits.
6.2.8OverviewtoHoldFunctions
ThefollowingtableliststhevariousHoldfunctionsalongwiththeirresetsandtimingvalues.Usethis
tableasaguideforprogrammingyoursensorandadjustingtheHoldtimes.
Pleasenote, the setting of somecommands isnot possibleby usingof the control panel,these
commandsareonlyavailable
bymeansofthesoftware.
Hold Function RESET by Peak Time Valley Time Threshold Hysteresis
Decay Rate
Protocol code
P F C XY XE
none none 000.0 000.0 -* -* -*
Peak Hold timer 000.1-299.9 000.0 000.0 -* 000.0
Peak Hold trigger 300.0** 000.0 000.0 -* 000.0
Hot objects moving on a production line
Standard output signal