Pride Mobility 1400 Mobility Aid User Manual


 
Go-Chair www.pridemobility.com 35
VIII. CARE AND MAINTENANCE
When to See An Authorized
Pride Provider for Service
The following symptoms could indicate
a serious problem with your Go-Chair.
If necessary, contact an authorized Pride
Provider. When calling, have the model
number, serial number, nature of the
problem, and the error code if available.
! Motor noise
! Frayed harnesses
! Cracked or broken connectors
! Uneven wear on any of the tires
! Jerky motion
! Pulling to one side
! Bent or broken wheel assemblies
! Does not power up
! Powers up, but does not move
Corrective Maintenance
If the battery condition meter does not
light up when you turn on the power:
! Check the harness connections.
Make sure they are tight.
! Check the circuit breaker. Reset it if
necessary.
! Check the battery connections.
If the above conditions prove normal,
you can load test the batteries with a
battery load tester. These testers are
available at automotive parts stores. Dis-
connect both batteries before load test-
ing and follow the directions that come
with the load tester. If either one of the
batteries fails the load test, replace both
of them. If your Go-Chair still does not
power up, contact an authorized Pride
Provider.
Figure 25. Go-Chair Battery Box