DISASSEMBL Y PROCEDURE
1.
B·tube
2.
F·tube
DISASSEMBL Y PROCEOURE
1.
B·tube
Release
screws
@x3,
used
for
attaching
main
body
(D
and
central
axle
C@D.
and
separate
the
binocular body
from
the
depression
prism
unit.
1)
Binocular body
For disassembly, refer
to
Fig.
2
and
3, taking the cautions
as
below:
To
dissolve
adhesive
No.
350, apply alcohol.
The
other adhesives, not dissolvable with alcohol,
is
to
be
removed
by
striking with a plastic
or
wooden hammer.
If
not removable, break the
glass
and
se
rape
off
the
adhesive.
e
To
separate
No.
2002, apply a drop
of
toluene.
2)
Depression prism unit
For disassembly, refer to
Fig.
1,
taking the cautions
as
below:
e
To
dissolve
adhesive
No.
350, apply alcohol.
No.
2001
being not dissolvable,
use
a plastic
or
wooden hammer. lf
not
removable, break the
glass
and scrape off the adhesive.
2. F-tube
Release
screws @
x3,
and
separate the binocular body
from
the depression
prism
unit.
1)
Binocular body
For disassembly,
foilow
the
sarne
procedure
as
for
B-tube.
2)
Depression prism unit
To disassemble the prism unit, refer to
Fig.
4, S and
6,
taking the cautions
as
below:
To
dissolve
adhesive
No.
350, apply alcohol.
No.
2001,
No.
921,
501
and 616 being not dissolvable with alcohol.
use
a plastic or wooden
hammer to strike
it out.
If
not removable, break the
glass
and scrape off
the
adhesive.
e For dissolving
No.
2002, apply a drop
of
toluene.
e
Take
off
only the upper
one
of roiler guides
@,
the lower
one
being
to
be
remained
as
fiducial.
-