R-4108.A
REPAIR ANO AOJUSTMENT
1. Failure of the Slide Glass Retainer 2. Slack
of
the
Slide Glass Carrier
3. Wobbling
of
the
Slide Glass Retainer and
the
Slide Glass Carrier
REPAIR
ANO
AOJUSTMENT
1.
Failure
of
the
Slide Glass Retainer
Cause: (1)
The
yielded
or
disengaged
sp1ral
spring @.
(2) Looseness of the
small
t1at
head
screw@.
< Repair and Adjustment >
(!)
Exchange
the
yielded spiral spring @
or
hook the disengaged spring
to
the
pin
hole.
Procedure j
i)
Disassemble the retamer assembly following the procedure
2),
3)
(P.
5).
ü)
If
the spring
is
yielded replace
it
with a spare, and if
the
spring
is
disengaged, reposition
it
correctly
foilowing
the reassembling proced.ure (P.
7).
ili)
Check the function of repaired retainer.
(2)
If
the small ílat
head
screw
@
is
loosened fasten
it
firm
and check the retainer
function.
2.
Slack
of
the
Slide
Gim
Carrier
Cause: Looseness
of
the
screws
@
< Repair
and
Adjustment >
Fasten
firm
the
screws
@ .
Make
sure
that
no
slack
is
found.
3. Wobbling
of
the
Slide Glass Retainer and
the
Slide
Gim
Carrier.
Wobbling in X direction: Refer
to
the
following chapter 4.
Wobbling
in
Y direction:
Refer
to the following chapter
5.
-
12
-