Nikon 50i and E100 Microscope & Magnifier User Manual


 
R-4108.A
REPAIR ANO· ADJUSTMENT
5.
Wobbling in Y (to
and
fro) Movement
(1
)-b)
If
any
dent
is
found on the guide rail @ , the
roUers@
or
if
the
retainer(2)
is
deformed replace them
wi
th spares.
Procadure
i) Detach the slide holder assembly from the stage releasing the screws @ x
2.
ii)
Remove the
rack@
releasing the screws
@x2.
iü) Remove the handle assembly releasing the screws @ x 2
of
the pinion
cover(V.
iv) Releasing the screws
@)
x 4 separa te the X guide block
(D
from the Y guide
blocks
@ , @ . Then, the X movement block
(D,
the
roUers{0
and the
retainers
@ , @
will
be made free.
v)
Releasing the
screws@xJ,
remove either @ or
3B
from(i)
leaving
another
one fixed on
G)
to make the basis for the reassembling. Then, the base
plate
(D,
the
rollers@
and the
retainers(Z)
become free.
vi)
If
any
dent
is
found on the Y guide rails @ , the
rollers@
or
if
the retainer
(2)
is
found
defonned
replace them with spares.
vii) Follow the reassembling procedure.
viii)
Check wobbling in the Y movement.
-
16
-