Mintek LV Pacemaker User Manual


 
Stratos LV/LV-T Technical Manual 163
Appendix A
Known Software Anomalies
Anomaly Possible Effect on Patient
or Implant Procedure
General Programmer Issues
If a long patient name is entered
without blank spaces, it may be
cut off in the printout under the
section 'Patient Data'
The last letters of the patient
name may be missing from
the printout. However, the
patient name is not a therapy-
relevant parameter.
Pacemaker detection disturbed
after interrogation under EMI
conditions
If this occurs, the programmer
must be re-started, so that the
Stratos cannot be
communicated with for about
1 minute. The follow-up
process is delayed by this
time.
Display test results on
programmer screen and printout
is inconsistent
Limited clarity and
consistency in the display of
measurement values from
follow-up testing.
During transfer of follow-up
information from Stratos CRT-Ps
to the GE/CARDDAS system, the
following data are not displayed:
the programmed data and
follow-up data of the left
ventricular channel
lead data of all channels in
Stratos
Missing programing
information in the exported
data may cause user
confusion
Interrogation data after Auto
Implant Detect (1.5 h after lead
attach) may show an incorrect,
superfluous entry: "AF episodes
detected >48h" in the Event List.
AF statistics are always correctly
displayed.
Display/printout of incorrect
diagnostic information may
cause user confusion or
incorrect diagnosis