Mintek LV Pacemaker User Manual


 
Stratos LV/LV-T Technical Manual 159
11.7 Default Programs
Stratos LV / LV-T
Parameter/
Function
Factory settings/
Standard Program
Safe Program
Mode DDD VVI
VV
Synchronization
BiV RV RV-T -
AV delay 150 ms at 60 ppm 120ms at 130
ppm
First Chamber
Paced
RV -
VV delay after
pace
5 ms -
VV delay after
sense
5 ms -
Pulse
Amplitude
3.6 V (A); 3.6 V
(RV); 3.6 V (LV)
4.8 V
Sensing A/RV UNIP UNIP (RV)
Refractory
Period
Auto (A); 250 ms (V) A: N/A
V: 300 ms
11.8 Materials in Contact with Human
Tissue
Housing: Titanium
Connector receptacle: Epoxy resin
Sealing Plugs: Silicone Rubber
11.9 Electrical Data/Battery
N
OTE:
At 37º C, with pacing impedance of 500 Ohms.