Mintek LV Pacemaker User Manual


 
148 Stratos LV/LV-T Technical Manual
Table 32 shows the mean
*
expected time intervals (in months)
from ERI to EOS at standard program for Stratos CRT-Ps. All
service intervals, including the above-cited nominal longevity,
are based on the battery discharge behavior and the hybrid
circuit properties including current consumption and replacement
indicator. The statistical calculations are based on 500 ohm lead
impedances, 100% pacing, and data supplied by the battery
manufacturer.
Table 32: Remaining Expected Service time (ERI to EOS)
Pacing Program Stratos LV, (months)
Standard
, Mean 8
Program with high pulse energy
8
*
50% of all pacemakers reach or exceed the given value
Standard: DDDR, 60 ppm, 3.6 V, 0.4 ms
High: DDDR, 90 ppm, 4.8 V, 1.0 ms