110 Stratos LV/LV-T Technical Manual
2.18.6 General Statistical Information
The Stratos CRT-P’s statistics modes are always in
operation and cannot be selected OFF.
The counters within the statistic features do not operate
when a magnet is applied to the CRT-Ps
The counters within the statistic features are reset each
time the Stratos CRT-Ps are permanently programmed.
Event counters are displayed as bar charts showing the
event totals expressed as a percentage.
Histograms count how often events occur in different
time or rate intervals (for example, how many events
occurred in a 160 to 169 ppm range).
Trends represent a certain number of events at a fixed
point in time (e.g., rates). The trends are plotted as
points that are joined together by a curve. In the
Stratos CRT-Ps, the trends of the three different
channels are identified with different colors.
NOTE:
When a magnet is applied, Stratos CRT-Ps can only
continue recording diagnostic data when in the “synchronous
magnet mode or during the first 10 cycles of the “auto”
magnet mode.
2.19 Interrogating and/or Starting
Statistics
The recorded diagnostic data (saved data contents of the
pacemaker) are always read (transmitted through interrogation)
at the beginning of the follow-up and saved to the programmer.
This allows relevant data to be displayed on the programmer at
anytime.
All statistics can be reset with the “Clear Statistics” function and
the “Transmit” function in the “Parameter” selection menu. The
start time and duration are automatically saved for all statistics.