Mintek LV Pacemaker User Manual


 
154 Stratos LV/LV-T Technical Manual
Intervention Rate
100… (10)…250 ppm
X-out-of-8 Activation Criterion
3… (1)…8
Z-out-of-8 Termination Criterion
3… (1)…8
DDI(R) Basic Rate
32… (1)…88… (2)…122… (3)…140… (5)…180 ppm
Overdrive Modes:
Off; On
Available in the modes DDD, DDTA, AAI, AAT with and without
rate adaptation
Maximum Overdrive Pacing Rate
90… (5)…160 ppm
Overdrive Pacing Increment (Rate Increase)
2… (2)…10 ppm
Overpacing Level (Rate Drop After)
1… (1) …32 cycles
Post-AES Pacing
Off; On
AES Increment
5… (5)…40 ppm
AES Prematurity
5…50%
VES discrimination after As
OFF; 250… (50)…450 ms
Rate Fading
Off; On