Lifescan Diabetes Monitoring System Blood Glucose Meter User Manual


 
MESSAGE PROBLEM WHAT TO DO
clean
test
area
1. There is dirt, blood, or
lint on the test area.
2. Your hand or an object
covered the test area
while the meter was
turned on.
3. The meter was used in
very bright light.
The test strip was
inserted before the
word insert appeared
on the display.
1. Clean the test area and
test strip holder according
to instructions.
2. Repeat the test. Keep the
meters test area clear.
3. Move the meter away
from the light source.
Repeat the test with a
new test strip.
Repeat the test with a new
test strip. Wait for the word
insert before inserting a
test strip.
remove
strip
52