Lifescan blood glucose monitor Blood Glucose Meter User Manual


 
57
PROBLEM
1. There is dirt, blood, or lint
on the test area.
2. Your hand or an object cov-
ered the test area during testing.
3. The meter was used in very
bright light.
The test strip was inserted
before the word INSERT
appeared on the display.
If the word CONTRL appears
after a blood test, your blood
sample was too small, smeared,
or another drop was added after
the test began.
WHAT TO DO
1. Clean the test area according to
instructions.
2. Repeat the test. Keep test area
clear.
3. Move the meter away from the
light source. Repeat the test with a
new test strip.
Remove the strip. Wait for the word
INSERT before inserting test strip.
Repeat the test with a new test strip
and apply a larger drop of blood.
MESSAGE
REMOVE
STRIP
CLEAN
TEST
AREA
100 MG/DL or
5.6 MMOL/L
(for example)
CONTRL
8:01AM 6-01-00