Leica M695 Microscope & Magnifier User Manual


 
Inspection procedure
09/93 3691-DMF
Service Manual M695
Doc Code 636 718
2.1
Turn LAMP potentiometer on control unit from min. to max.
Reading on test box: LAMP REF.
min./max. 0.3......5v.
The brightness of the microscope lamp is continuously adjustable.
2.2
Activate the quick-change lamp mount.
The spare bulb is moved into the beam path, and lights up.
The orange indicator springs out.
(Push the orange indicator back into the rear wall)
(Fault location in lamp control: page 4-7)
3.1
Turn the ZOOM SPEED potentiometer from min. to max. and from max. to min.
Reading on test box: ZOOM SPEED
min./max 1.84...2.34Vl5V ±5%
The zoom speed is continuously adjustable.
3.2
Use foot-/handswitch to activate ZOOM UP.
Reading on test box: ZOOM UP
Low/High <0.2V/>4.75V
3.3
Use foot-/handswitch to activate ZOOM DN.
Reading on test box:
ZOOM DN
Low/High <0.2V/>4.75V
(Fault location in ZOOM control: page 4-8)
3.4
Mechanical zoom movement.
Zoom moves uniformly in both directions.