Service manual M695 Description swingarm MSV136/137
Doc code 636 718
09/93 3077-HWE
List of signals for swingarm MSV 136/137
Required values for standard two-metre swingarm cable
•
Unless otherwise indicated, set the lamp to minimum brightness, because
the values quoted are only valid under these conditions.
Power supply
SIGNAL
Voltage against PGND
Lamp min. brightness ->
Voltage against PGND
Lamp max. brightness ->
Voltage against SENSE
Lamp min. brightness ->
Voltage against SENSE
Lamp max. brightness ->
Voltage drop against logics level
Lamp min. brightness: ->
(Test box: Microscope
Setting: Focus up)
Voltage drop against logics level
Lamp max. brightness ->
(Test box: Microscope
Setting: Focus up)
VALUE
26V
20V
-50mV
-250mV
4.9V
4.2V
TOL.
+-5%
+-5%
-0.2v
-0.8V
6-1