Invacare X-Terminator Titanium Wheelchair User Manual


 
SECTION 5—WHEELS
Part No 1114851 41 Everyday and Sport Series Wheelchairs
Adjusting Rear Wheel Camber
NOTE:Forthisprocedure,refertoFIGURE 5.7.
Camber Tube 2000
Theaxletubemustbereplacedtochangethecamber.RefertoReplacingAxleTubeon
page 49.
A4 Camber System
NOTE:Performthisproceduretoonesideof
thewheelchairatatimeforeaseofadjustment.
1. Openthecamberclamp.Referto
Opening/ClosingCamberClampson
page 39.
2. Pullonerearwheelwithcamberinsert
outofaxletube.
3. Removetherearwheelfromthe
camberinsert.
4. Rotatethecamberinsert180°.
5. Reinstalltherearwheelintothecamber
insert.
FIGURE 5.7 Adjusting Rear Wheel Camber
- A4 Camber System
ƽ WARNING
NEVER position the camber inserts in the axle tube with more than 3-inches (12
indexing marks showing) of the camber insert outside of the axle tube. The camber
inserts will not be securely tightened in the axle tube. Possible injury to the user
and/or damage to the wheelchair may occur.
6. Positionthecamberinserttothedesiredposition.Makesurethereisnomorethan
3inches(12indexingmarks)ofthecamberinsertsoutsideoftheaxletube.
7. Slidetheindexingringonthecamberinsertuntilitisflushwiththecamberclamp.
NOTE:Beforeusingthewheelchair,makesurebothcamberinsertsaresettothesameindexing
mark.Thiswillmakesurethedistancebetweentherearwheelandthewheelchairisthesameon
bothsides.
ƽ WARNING
QUICK RELEASE LEVERS - Make sure the quick release levers are pointing
towards the rear of the wheelchair and are in the closed position before using the
wheelchair, otherwise personal injury or damage to the wheelchair may occur.
Rear View of Wheelchair
Camber
Clamp
Axle
Tube
Camber Insert -
Rotate 180°
Indexing Ring