SECTION 8—MAINTENANCE
Part No 1141431 35 Reliant 440
FIGURE 8.1 Replacing Electric Actuator
Maintaining the Base Adjustment
ƽ WARNING
This procedure MUST be performed by a qualified technician.
NOTE: For this procedure, refer to
FIGURE 8.2.
The base adjustment should not require
any attention other than:
1. Check the squareness of the legs when
in the closed position.
2. Place a 90° square on the inside of the
legs and base to determine the 90°
alignment.
3. Adjust the linkage rods until 90°
alignment is achieved.
FIGURE 8.2 Maintaining the Base
Adjustment
Actuator Power Cord*
Top Washers
Mast Mounting Bracket
Bottom Mounting Screw
Electric Actuator
Bottom Washer
Top Locknut
Top Mounting Screw
Lift Arm
NOTE: The two plastic caps covering the top locknut and the top mounting screw are not shown.
*NOTE: Actuator power cord shown disconnected from the control box.
Lift Arm
Mounting
Bracket
Control Box
Bottom Locknut
Bottom Washer
90° Square
Linkage
Rod