SECTION 7—TROUBLESHOOTING
Reliant 440 30 Part No 1141431
SECTION 7— TROUBLESHOOTING
NOTE: If problems are not remedied by the suggested means, please contact your dealer or
Invacare.
SYMPTOMS FAULTS SOLUTION
Stand-up lift feels loose. Mast/base joint loose.
Tie - rods are loose.
Tighten the mounting screws that secure the
mast to the base. Refer to Assembling the Patient
Lift on page 13.
Refer to Maintaining the Base Adjustment
on
page 35.
Casters/brakes noisy or
stiff.
Fluff or debris in bearings. Replace casters.
Electric actuator fails to
lift when button is
pressed.
Hand-control or actuator con-
nector loose.
Batteries low.
RED emergency stop button
pressed IN.
Battery not connected prop-
erly to control box.
The connecting terminals are
damaged.
Electric actuator in need of ser-
vice or load is too high.
Check connections.
Charge batteries. Refer to Charging the Battery
on page 19.
Rotate RED emergency stop button CLOCK-
WISE until it pops out.
Reconnect the battery to the control box. Refer
to Charging the Battery
on page 19.
Replace the battery pack. Refer to Charging the
Battery on page 19.
Refer to Replacing Electric Actuator on page 34.
Contact your Dealer.
Unusual noise from
actuator.
Actuator is worn or damaged
or spindle is bent.
Refer to Replacing Electric Actuator
on page 34.
Contact your Dealer.
Lift arms will not lower
in uppermost position.
Lift arms require a minimum
weight load to lower from the
uppermost position.
Pull down slightly on the lift arms.