IBM M/02-212 Scale User Manual


 
Brookfield Engineering Labs., Inc. Page 58 M/02-212
A2 Error messages
The R/S Rheometer is rather error resistive, when errors occur they are determined and the user is
informed on the LCD display.
The most frequent error massages are explained in this Chapter.
Range Error
Cause: User’s error at input. The user has tried to input a value that is less than the allowed
minimum value or is more than the allowed maximum value.
Message:
Example of error input of a preset value
What to do: Input the value again. Be sure that new value is inside of the allowed value range.
Printer Error
Cause: The printer is selected as output device and is not ready for printing.
a) Printer cable is not connected to the Rheometer
b) Paper out, printer not on line, other printer errors
Message:
Example: printer not ready
What to do: Check if the printer cable is connected to the Rheometer.
Check if paper is in the feed sheeder of the printer.
Check if the “Ready” LED lights on printer.
Remove the error cause at printer and push key at Rheometer. If you cannot find the
error cause, push key in order to break the action that is carrying out now.
Range error
End value:
min : 0.90
max : 1032.80
Error #1
Printer not ready!
<OK> cont. <ST> stop
Name of value which is out of ran
g
e
Allowed minimum value
Allowed maximum value
Cancel action
Try again
OK
ST