All free memory parts are marked as NEW. In order to avoid overwriting of already available programs,
please, select a NEW program to be edited.
After selection of the program with key you will be requested to select a measuring system.
Attention: the program has to be executed later with the same measuring system, otherwise senseless
measurement results will be calculated.
Select Meas. system:
3) CC45 >
4) CC25
5) CC14 >
After selection of the measurement system, input of number of steps requested (=number of slopes and
planes of the function).
Number of steps can be from 1 to maximum 10 steps.
In our example we need 3 steps. So, number “01” must be changed to “03”. (Input of numbers - see
Chapter 6.4).
If you try to input numbers <1 or >10, an error message “Range-Error” will appear.
After successful input of number of steps, selection of kind of measurement is carried out:
Possible types of measurement Range of parameters preset
Shear rate D [s
-1
] Depending on measuring system
Speed n [rpm] 0.7 ... 800 rpm
Shear stress Tau [Pa] Depending on measuring system
Torque M [‰] 0 ... 999 promille
So selectent (in our example) “D[1/s]” and press key. Now, you have to define each step.
Select PROG to edit
1) Prog xyz
2) Test
3) NEW >
Nr. of steps : 01
Select Input-Mode:
-D[1/s]
-n[rpm]
-Tau[Pa] >
Input number of
steps 1...10
Our example: 03
All free (non-defined) programs
are marked as NEW
Available programs
contain a name
OK
OK