IBM M/02-212 Scale User Manual


 
Brookfield Engineering Labs., Inc. Page 33 M/02-212
At this point you either can start a measurement by pressing key, or return back to Main menue
using key.
In this example, we see that the measurement data will be saved in Memory only. No other output
device is defined. Output devices are printer or serial interface RS232 of the Rheometer. Selection of
these devices is described in Chapter ”Configuration -> Set output mode”.
In case, the Memory is full with measurement data, you have to cancel function Run Single, print out the
measurement data from the memory or send data to a PC via serial interface (see Chapter “Utilities
Print Memory”) and then delete measurement data from the Memory (see Chapter “Utilities Delete
Memory”).
However, you can start a measurement even with full Memory, in this case measurement results will be
shown only at LCD display and will be lost after completion of the measurement (because no output
device archives measured data).
If you start the measurement, the instrument shows:
This message will be displayed until the first measurement point is calculated.
First Measure point is displayed like this:
If torque is less than 10 ‰
measurement results are often of low precision. In this case, preset values
are to be changed in such a manner, that the torque is higher than 10 ‰.
If the display field for the torque indicates: “M low!”, these values are outside the range of resolution of
the Rheometer.
If the temperature is not to displayed, the measuring sensor is not connected.
At printout, the temperature is printed as 1000.0°C!
The display is updated on every new measure point. Current measurement can be canceled at any time
by pushing key.
Output of MPs to:
-No output device
-Memory
<^> Menu <ST> Start
Prog. Running...
Wait for 1.MP
#1
100.5 ‰ 000.90 1/s
Eta: 1.728 Pas
Tau: 1.572 Pa
04s 20.7 °C #01
Actual value of preset
p
arameter
Measurement result
Step (in Single meas.
always 01)
Tor
q
ue
Duration
Temperature
ST
ST