Brookfield Engineering Labs., Inc. Page 12 M/02-212
Please read information concerning installation of the computer system in this Operation Manual.
III.3 Mounting of temperature controlled measuring devices
In this chapter you get information concerning mounting and attachment of the following accessories
components:
FTK-CC temperature control device for use of cylinder measuring system in the
temperature range -10°C...+90°C (liquid’s temperature control)
KE cooling device in conjunction with FTK-CC temperature control device or ME-CP/PP
measuring device to rise the temperature range up to -20°C...+180°C
ME-CP/PP measuring device for use of cone/plate and plate/plate measuring system in the
temperature range -10°C...+90°C (liquid’s temperature control)
III.3.1 Mounting of temperature control device FTK-CC with and
without cooling device KE
As temperature control device is available optionally:
- FTK-CC
for use of cylinder measuring system in the temperature range -10°C...+90°C (liquid’s temperature
control).
The temperature control device “FTK-CC” should be used only in the temperature
range -10°C ...+90°C.
The cooling device “KE” is necessary additionally for temperature range of -
20°C...+180°C!
hose connection of flow output
Pt 100
thermostating chamber
hose connection of flow input
measuring cup screw fitting
thread
j
oint for mountin
g
flan
g
e
Fig.: 4.: Temperature control device FTK-CC
Mounting of the temperature control device “FTK-CC”
- Turn off the R/S Rheometer with the mains switch “POWER” at the instrument back side.
- When using cooling device KE mount cooling device first (see chapter 3.3.3)
- Set the “FTK-CC” from below on the mounting flange of R/S Rheometer and tighten the thread.
- Fix the hoses of liquid circulation thermostat at the “ FTK-CC ” as described below.
- Connect the FTK-CC´s built-in Pt100 cable with the connector “PT100” at the instrument back side
of the R/S Rheometer.