Fluke 1522 Thermometer User Manual


 
Description Command Example
Response
Format
Response
Example Range
Read probe wires w[ires] wi wi: 9 wi: 4
Set probe wires w[ires]=2/3/4 wi=4 2,3,4
Read probe calibration date pc[al] pcal pc: yyyy-mm-dd pc: 1999-03-28
Set probe calibration date pc[al]=yyyy-mm-dd pc=1999-03-28 1971-01-01 to
2099-12-31
Read probe due date pd[ue] pdue pd: yyyy-mm-dd pd: 1999-03-28
Set probe due date pd[ue]=yyyy-mm-dd pd=1999-03-28 1971-01-01 to
2099-12-31
Read filter fi[lter] fi fi: 99.999999 fi: 0.5
Set filter fi[lter]=n fi=0 0.0 to 60.0
Read low range calibration *c1 *c1 *c1: 9.999 *c1: -0.0121
Set low range calibration *c1=n *c1=-0.012 –1.999 to
1.999
Read high range calibration *c2 *c2 *c2: 9999.9 *c2: -0.9
Set high range calibration *c2=n *c2=-0.9 –1980.0 to
1980.0
Read meter calibration date mc[al] mcal mc: yyyy-mm-dd mc: 1999-03-28
Set meter calibration date mc[al]=yyyy-mm-dd mc=1999-03-28 1971-01-01 to
2099-12-31
Read meter due date md[ue] mdue md: yyyy-mm-dd md: 1999-03-28
Set meter due date md[ue]=yyyy-mm-dd md=1999-03-28 1971-01-01 to
2099-12-31
Legend: [] Optional Command data
{} Returns either information
n Numeric data supplied by user
9 Numeric data returned to user
x Character data returned to user
51
8 Communications Interface
Communication Command List