Description Command Example
Response
Format
Response
Example Range
Set autolog mode on/off .
Note: Exit Autolog menu
to use this command. After
alog =ON is sent, the buttons
are disabled until the command
alog=OFF is sent.
alog=[ON/OFF] alog=ON ON or OFF
Clear autolog memory aclr aclr
Initiate download of autologged
data.
Note: Read Section
11.2 on downloading
autologged data before at
-
tempting to use this
command.
daut daut
Read autolog interval aint aint int: 99 {secs or min} int: 60 {secs or min}
Set autolog interval
Note: See Section 7.4.2
for specifics on log interval.
aint=n aint=30 1, 5, 10, 10,
15, 30, 60,
120, 300, 600,
900, 1800,
3600
Read delta(x) measurement dx dx dx: 999.999
{C,O,F,R,orK}
dx: -0.118 C
Set delta(x) reference to pres-
ent measurement
sd[x] sd
Readunit u[nit] u u:{C,O,F,R,orK} u:C
Set unit u[nit]=C/O/F/K/R u=C C, O, F, K, R
Read serial sample rate sa[mple] sa sa: 99999 sa: 10
Set serial sample rate sa[mple]=
n
sa=60 0 to 10,000
Read resolution digits re[solution] res res: 9 res: 3
Set resolution digits re[solution]=
n
res=0 1 to 3
Read time ti[me] time ti: hh:mm:ss ti: 16:23:45
Set time
Note: You must set
hours, minutes, and seconds
for the time to be set
correctly.
ti[me]=hh:mm:ss ti=08:15:00 00:00:00 to
23:59:59
Read identification information *i[dn?] *idn? <manufacturer>,
<model>,
<serial number>,
<firmware version>
Hart,
1522
95001,
1.24
Read model and firmware
version
*v[ersion] *ver ver.<model>,
<firmware version>
ver.1522,
1.24