E8 ATEX Flameproof and Dust
Certificate: KEMA00ATEX2013X; Baseefa11ATEX0275X
Standards: EN60079-0:2012, EN60079-1:2007, EN60079-26:2007, EN60079-31:2009
Markings: II 1/2 G Ex d IIC T6/T5 Ga/Gb, T6(-50 °C T
a
+65 °C),
T5(-50 °C T
a
+80 °C);
II 1 D Ex ta IIIC T95 °C T
500
105 °C Da (-20 °C T
a
+85 °C)
Special Conditions for Safe Use (X):
1. This device contains a thin wall diaphragm. Installation, maintenance and use shall
take into account the environmental conditions to which the diaphragm will be
subjected. The manufacturer's instructions for installation and maintenance shall
be followed in detail to assure safety during its expected lifetime.
2. For information on the dimensions of the flameproof joints the manufacturer shall
be contacted.
3. The capacitance of the wrap around label, being 1.6 nF, exceed the limit in Table 9
of EN60079-0. The user shall determine suitability for the specific application.
4. Some variants of the equipment have reduced markings on the nameplate. Refer to
the Certificate for full equipment marking.
I1 ATEX Intrinsic Safety and Dust
Certificate: BAS97ATEX1089X; Baseefa11ATEX0275X
Standards: EN60079-0:2012, EN60079-11:2012, EN60079-31:2009
Markings: II 1 G Ex ia IIC T5/T4 Ga, T5(-60 °C T
a
+40 °C), T4(-60 °C T
a
+70 °C)
II 1 D Ex ta IIIC T95 °C T
500
105 °C Da (-20 °C T
a
+85 °C)
Special Conditions for Safe Use (X):
1. The apparatus is not capable of withstanding the 500 V insulation test required by
EN60079-11. This must be taken into account when installing the apparatus.
2. The enclosure may be made of aluminum alloy and given a protective polyurethane
paint finish; however care should be taken to protect it from impact or abrasion if
located in Zone 0.
3. Some variants of the equipment have reduced markings on the nameplate. Refer to
the Certificate for full equipment marking.
Table 6. Process Temperature
Temperature class Process temperature
T6 -50 °C to +65 °C
T5 -50 °C to +80 °C
Table 7. Input Parameters
HART
Voltage U
i
30 V
Current I
i
200 mA
Power P
i
0.9 W
Capacitance C
i
0.012 μF
Inductance L
i
0 mH