Emerson Process Management 3051 Pacemaker User Manual


 
Quick Start Guide
15
February 2015
Trimming with configuration buttons
A zero trim is to be performed using one of the three possible sets of external
configuration buttons located under the top tag.
To access the configuration buttons, loosen the screw and slide the tag on the top
of the transmitter. Confirm the functionality using Figure 10.
Figure 10. External Configuration Buttons
Use the following procedures to perform a Zero Trim:
Perform trim with LOI (option M4)
1. Set the transmitter pressure.
2. See Figure 9 on page 13 for the operating menu.
a. Perform an analog zero trim by selecting Rerange.
b. Perform a digital zero trim by selecting Zero Trim.
Perform trim with analog zero and span (option D4)
1. Set the transmitter pressure.
2. Press and hold the zero button for two seconds to perform an analog zero
trim.
Perform trim with digital zero (option DZ)
1. Set the transmitter pressure.
2. Press and hold the zero button for two seconds to perform a digital zero trim.
A. LOI
B. Analog Zero and Span
C. Digital Zero
D. Configuration Buttons
A
BC
D