Emerson 3000 Oxygen Equipment User Manual


 
Instruction Bulletin
106-300NE Rev. 3.4
May 2000
3-2 Setup Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
NOTE
The voltage given above is for an am-
bient (machinery space) temperature
of 77°F (25°C). For each degree of am-
bient temperature above or below 77°F
(25°C), add or subtract 0.242 mV from
the nominal. Example; at 87°F (31°C),
the nominal voltage of -322.3 should
be increased (made less negative) by
10 x 0.242 or 2.42 mV, making the ad-
justed nominal -319.9.
f. Remove voltmeter leads.
g. Remove jumper wire.
3-3 MODEL 218A ELECTRONICS SETUP
Before beginning operation, it is important that
the probe heater set point of the existing elec-
tronics be changed to support the World Class
3000 probe. To convert the Model 218A Digital
Electronics Package for use with the World
Class 3000 Probe and Heater Power Supply,
an EPROM change is necessary. Remove Main
PCB and check back of board to identify unit as
G02 or G04. The replacement EPROM needed
is as identified below:
G02 G04
United States 1M03192G01 1M02982G01
United Kingdom 1M03192G02 1M02982G02
Germany 1M03192G03 1M02982G03
France 1M03192G04 1M02982G04
Italy 1M03192G05 1M02982G05
NOTE
The replacement EPROM when using a
multiprobe averager unit is
1M02982G10.
To replace the EPROM, proceed as follows:
a. Shut off and lock out power to the electron-
ics package.
b. Open electronics enclosure.
c. On the Main PCB, Figure 3-2, locate and
remove old EPROM.
d. Replace with new EPROM.
e. Close electronics enclosure and power up
system.
Figure 3-2. Main PCB (Model 218A) EPROM Replacement