Emerson 3000 Oxygen Equipment User Manual


 
Instruction Bulletin
106-300NE Rev. 3.4
May 2000
Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11
World Class 3000
3D3 080G REV
JM1
JM2
JM3
JM4
JM5
JM6
JM7
JM8
22540005
NOTE:
FUSE
FUSES
ALL FUSES ARE 5A @ 250 VAC,
ANTI-SURGE, CASE SIZE
5 X 20 MM, TYPE T TO IEC127,
SCHURTER.
Figure 2-7. Jumpers on HPS Mother Board
NOTE
!
Upon completing installation, make sure that the probe is turned on and operating
prior to firing up the combustion process. Damage can result from having a cold
probe exposed to the process gases.
During outages, and if possible, leave all probes running to prevent condensation and
premature aging from thermal cycling.
If the ducts will be washed down during outage, MAKE SURE to power down the
probes and remove them from the wash area.
2