A&D HV-60KV-WP Scale User Manual


 
Page 46 HV-WP/HW-WP Series
15. The Function Table
Item Display Meaning and purpose
The details of the full/dribble batch function (f6 9)
f10 0 #
Reaching final value and pressing the F switch, the current
process is finished.
A
A
f10 1
Reaching final value and displaying the stability mark, the current
process is finished automatically.
A
A
f10 2
Pressing the SET switch, the scale automatically tares and starts
the full/dribble batch process. Reaching final value and pressing
the F switch, the current process is finished.
A
A
A
Full/Dribble
batch sub-
func-tion
f10 3
Pressing the SET switch, the scale automatically tares and starts
the full/dribble batch process. Reaching final value and
displaying the stability mark, the current process is finished.
A
A
A
Address for RS-422 / RS-485 for option OP-04
Address
f11 1 #
f11 99
Use This address can be set from 01 to 99. It is possible for a
computer to control the scale assigned an address.
(RS-232C should be set to "00".)
The function to hold the current display value. When the value
becomes nearly-zero and the weighing value changes more than
25% +30 digits, hold display is canceled.
f12 0 # Not used
f12 1
The hold function is ON or OFF alternately by the F switch.
Hold
f12 2
Displaying the stability mark, the display is held.
Selection of readability and response by averaging the weighing value.
f13 0 # Normal weighing.
f13 1
Scale for a person’s weight.
Averaging
f13 2
Animal weighing.
Selects a storable minimum unit mass in the counting mode.
f14 0 # Stores a unit mass in the unit of a digit.
f14 1
Stores a unit mass in the unit of one eighth digit.
Precision
of unit
mass
f14 2
Stores a unit mass, if the total of the sample mass is greater than
5 digits.
Refer to "16.1. Data Format".
f15 0 # Format 1. (A&D general format for scales, balances)
Format
f15 1
Format 2. (Older HV-A/ HW-A format)
f16
Not used.
The "#" is factory settings
The "nearly-zero" is within ±4 digits from zero point in the unit of kg.
The "digit", a unit of display, is equivalent to the minimum measurable mass.