A&D HV-60KV-WP Scale User Manual


 
Page 42 HV-WP/HW-WP Series
14. Calibration (Adjusting Scale)
14.2.3.
14.2.3.14.2.3.
14.2.3. Calibration of the Zero Point
Calibration of the Zero PointCalibration of the Zero Point
Calibration of the Zero Point
Step 8 Press and hold the CAL switch to enter
the calibration mode after displaying
normal weighing for 30 minutes.
Then Cal 0 is displayed
Step 9 With nothing on the weighing pan, press
the ENTER switch while the stable mark is
displayed. The scale stores the current
condition as the zero point.
Step10 The scale displays 5pn1 for several
seconds. Finishing the calibration mode
at this stage, proceed to step 14.
14.2.4.
14.2.4.14.2.4.
14.2.4. Span Calibration
Span CalibrationSpan Calibration
Span Calibration
Step11 Set the value of the calibration mass
using the following switches. (This initial
value is according to each product.)
switch Selecting the number of a
figure.
switch Selecting a figure.
Step12 Place the mass on the weighing pan
which was set at step 11, press the
ENTER switch while the stable mark is
displayed. The scale then calculates the
span and stores it.
Step13 The scale displays end at the finish.
Remove the mass from the weighing pan.
Step14 Press the CAL switch to return to the normal
weighing mode.