HW-G Instruction Manual Page 47
HW-G OZIM v1a
Item Display Meaning and purpose
The function to hold the current display value. When the
value becomes near-zero and the weighing value changes
more than 25% +30 digits, hold display is cancelled.
Hold f12 0 # Not used
f12 1 The hold function is ON or OFF alternately by the F switch.
f12 2 Displaying the stability mark, the display is held.
Selection of readability and response by averaging the weighing
value.
Averaging f13 0 # Normal weighing.
f13 1 Scale for a person’s weight.
f13 2 Animal weighing.
Selects a storable minimum unit mass in the counting mode.
Precision f14 0 # Stores a unit mass using the unit of a digit.
of f14 1 Stores a unit mass in the unit of one eighth digit.
unit mass f14 2 Stores a unit mass, if the total of sample mass is heavier
than 5 digits.
Refer to “16.1. Data Format”.
Data f15 0 # Format 1. (A&D general format for scales, balances)
Format f15 1 Format 2. (Older HV-A/ HW-A format)
It is necessary to use option OP-06 for type V.
Use Print out of date. Caution Date is not updated.
Print out of the management number for the scale.
Date f16 No print out.
000000
The relation between printout and display is as follows:
other Display 990721
Printout 99/07/21