A&D HG-60KGV Scale User Manual


 
HW-G Instruction Manual Page 19
HW-G OZIM v1a
7.2 Tare (and Net Display)
The “Tare” is used to cancel the mass of a container, receptacle, case, bag, etc.
which is put on the pan to contain the item to be weighed.
Caution
The tare reduces the available weighing range.
The current tare value is reset by pressing the ZERO switch when the pan is
empty, or by turning the scale off. (Reset value is zero.)
7.2.1 Semi-Automatic Tare (Input by Weighing)
Step 1 Put the container onto the pan.
Step 2 Press the TARE switch. The display becomes zero and the net mark
is displayed.
Step 3 It is now possible to put something into the container and to read its
net display.
Step 4 Remove all items from the pan.
7.2.2 Preset Tare (Digital Input of Known Tare)
Step 1 Press and hold the SET switch and press the TARE switch.
Then the blank or stored tare value is displayed. This blank display
means that the tare value is zero (reset value) and blinks.
Step 2 Set the preset tare value by using the following switches.
^ switch selecting the value of a digit.
< switch selecting a digit.
Step 3 Press the ENTER switch to store the new preset tare value.
The scale displays a net value i.e. the tare value subtracted from the
gross weight value.
Step 4 It is now possible to put something onto the pan and to read its net
weight.
Step 5 Remove all items from the pan.