HW-G Instruction Manual Page 45
HW-G OZIM v1a
Item Display Meaning and purpose ?
Comparator When the display value becomes stable after pressing the
SET switch, the scale compares the display value. It does not
f6 4 compare on an unstable condition. If the SET switch is pressed
again, the scale stops the comparison.
When the display value becomes stable, the scale compares
f6 5 the display value. It does not compare on an unstable
condition.
When the display value becomes stable and not near-zero
after pressing the SET switch, the scale compares the display
f6 6 value. It does not compare on an unstable condition. If the SET
switch is pressed again, the scale stops the comparison.
f6 7 When the display value becomes stable when not near-zero,
the scale compares the value.
f6 8 Simple batch function. (Refer to section 13.)
f6 9 Full/dribble batch function. (Refer to section 12.)
Buzzer
The condition of the buzzer on option OP-03 by comparator
function or full/dribble batch function.
f7 0 # No buzzer.
f7 1 The buzzer sounds at LO.
f7 2 The buzzer sounds at OK.
f7 3 The buzzer sounds at LO and OK.
f7 4 The buzzer sounds at HI.
f7 5 The buzzer sounds at LO and HI.
f7 6 The buzzer sounds at OK and HI.
f7 7 The buzzer sounds at LO, OK and HI.
f7 8 The buzzer sounds at finishing the full/dribble batch process.
Accumulator
The condition of the accumulator
f8 0 # Accumulator is not used.
The scale accumulates the data, if the F switch is pressed,
f8 1 when the display becomes a positive stable value above
near-zero. Next accumulation can be performed after the
display becomes near-zero or a negative value.
The scale accumulates the data, if the F switch is pressed,
f8 2 when the display becomes a stable value outside near-zero.
Next accumulation can be performed after the display be
comes near-zero.
When the display becomes a positive stable value, the scale
f8 3 accumulates the data automatically. Next accumulation can
be performed after the display becomes near-zero or a
negative value.