Teledyne Analytical Instruments 6
Remote Span/Zero Digital inputs allow external control
of analyzer calibration.
Calibration Contact To notify external equipment that
instrument is being calibrated and readings are not
monitoring sample.
Range ID Contacts Four separate, dedicated, range relay
contacts. Low, Medium, High, Cal.
Network I/O Serial digital communications for local
network access. For future expansion. Not implemented
at this printing.
Optional:
Calibration Gas Ports (Auto Cal Option) Separate
fittings for zero, span and sample gas input, and internal
valves for automatically switching the gases.
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3000TA is accurate to 2-3 %. Accordingly, internally
scheduled calibrations can vary 2-3 % per day.