Performan c e Verification
The following procedure verifies the warranted electrical characteris-
tics of the High Voltage Differential Probe. Table 7 itemizes the
equipment required, provides an example or part number of the
equipment, and explains the purpose of the equipment.
Table 7: List of equipment required
Description
Minimum
requirements
Example or part
number Purpose
Test oscilloscope Bandwidth: 400 MHz
TEKPROBE interface
vertical accuracy:
1.5%
TDS460A,
TDS744A, or
TDS784A
Display probe output
Standard ampli-
tude generator
Leveled sine wave
generator
DC voltage source
Amplitude accuracy:
0.75%
Rise time: 3ns
Fluke/Wavetek
9100 with oscillo-
scope option 250
Check probe attenua-
tion, bandwidth, and
common-mode rejec-
tion ratio
Plunger clamps
(2 required)
Use clamps included in
accessory kit
Accessory kit
020-2106-00
Connection to binding
post
Modified BNC
adapter
BNC-male-to-dual
binding post
103-0035-00 Interconnection be-
tween probe and gen-
erator. (See Figure 11
on page 28).
BNC adapter BNC-female-to-dual
banana
103-0090-00 Interconnection be-
tween probe and gen-
erator
Coaxial cables
(2 required)
36 in (0.9144 m),
precision 50
012-0482-00 Interconnection
between oscilloscope
and generator
Terminator
50 precision feed
through
011-0049-01 Termination between
probe and generator
during bandwidth mea-
surements