Tektronix 070-9472-02 Thermometer User Manual


 
Specifications
16
P5205 Instruction Manual
Table 2: Warranted Electrical Characteristics (Cont.)
Maximum Operating Input Voltage 500X differential: ± 1.3 kV (DC + peak AC)
500X common mode: ± 1kV
RMS
CAT II
50X differential: ± 130 V (DC + peak AC)
50X common mode: ± 1kV
RMS
CAT II
Maximum Nondestructive Input Voltage 500X and 50X differential: ± 1.5 kV (DC +
peak AC)
500X and 50X common mode: ± 1.3 kV
RMS
CAT II
Gain Accuracy
± 3% at 20-30_ C, <70% RH
Temperature
Operating: 0 to 40_ C
Nonoperating: -30 to +70_ C
Humidity
Operating: <85% RH at or below +35_ C
Nonoperating: <85% RH at or below
+60_ C
Table 3: Certifications and Compliances
EC Declaration of
Conformity -- Low
Voltage
Compliance was demonstrated to the following specification as
listed in the Official Journal of the European Communities:
Low Voltage Directive 73/23/EEC, as amended by 93/68/EEC:
EN 61010-1/A2:1995
Safety requirements for electrical equipment for
measurement, control, and laboratory use
EN 61010-2-031:1994
Particular requirements for hand-held probe assemblies for
electrical measurement and test equipment
Approvals
UL3111-1 - Standard for electrical measuring and test equipment
IEC 10106-2-031 - Particular requirements for hand-held probe
assemblies for electrical measurement and test
CAN/CSA-C22.2 No. 1010.1-92 and CAN/CSA-C22.2 No.
1010.2.031-94 - Safety requirements for electrical equipment for
measurement, control, and laboratory use