ResMed AutoSet CS Sleep Apnea Machine User Manual


 
English
7
SD card
An SD card has been supplied to gather therapy data from your device and provide settings updates
from your clinician. When instructed to do so, disconnect the device from the power outlet, remove your
SD card, insert it in the protective folder and send it to your clinician. For more information on removing
and inserting your card refer to the S9 SD Card Protective Folder provided with your device. Please
retain the S9 SD Card Protective Folder for future use.
Technical specifications
Operating pressure range 4 to 20 cm H
2
O (CPAP); 4 to 25 cm H
2
O (VAuto, ASV);
2 to 25 cm H
2
O (S, ST)
Maximum single fault steady state
pressure
30 cmH
2
O – if pressure exceeded for > 6 secs;
40 cmH
2
O – if pressure exceeded for > 1 sec
Pressure measurement tolerance ±0.5 cm H
2
O ± 4% of the measured reading
Flow measurement tolerance ±6 L/min or 10% of reading, whichever is greater
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level 24 dBA as measured according to ISO 17510-1:2002
26 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
Sound power level 34 dBA with uncertainty of 2 dBA as measured according to
EN ISO 17510-1:2009
Nominal dimensions (L x W x H) 153 mm x 140 mm x 86 mm
Weight 835 g
90W power supply unit Input range 100–240V, 50–60Hz,
Nominal for aircraft use 115V, 400Hz
Typical power consumption 70W (80VA)
Maximum power consumption 110W (120VA)
30W power supply unit Input range 100–240V, 50–60Hz,
Nominal for aircraft use 115V, 400Hz
Typical power consumption 20W (40VA)
Maximum power consumption 36W (75VA)
90W DC/DC converter Nominal inputs 12V, 24V
Typical power consumption 70W
Maximum power consumption 110W