ResMed AutoSet CS Sleep Apnea Machine User Manual


 
English
6
Displays error message: Check ResMed 30/90W Power Supply Unit and fully insert the
connector
The DC plug is partially inserted into the
back of the device or inserted too
slowly.
Fully insert the DC plug.
A non-ResMed power supply unit is
connected to the device.
Remove the power supply unit and replace with a ResMed
power supply unit.
The power supply unit is being covered
by bedding.
Make sure that the power supply unit is free from bedding,
clothes or other objects that could cover it.
Displays error message: Tube blocked, please check your tube
Air tubing is blocked. Check your air tubing and remove any blockages.
Disconnect the power cord and then reconnect it to restart
the device.
Displays error message: High leak, please check system setup and all connections
There is excessive leak. Adjust position of mask and headgear.
Air tubing is not connected properly. Connect firmly at both
ends.
The following message is displayed on the LCD after you try to update settings or copy data to
the SD card: Card error, please remove SD card and contact service provider
SD card is not inserted correctly. Ensure that the SD card is inserted correctly.
You may have removed the SD card
before settings were copied to the
device.
Reinsert the SD card and wait for the home screen or the
Settings updated successfully, press any key”
message to appear on the LCD.
Note: This message only appears once. If you re-insert the
SD card after you have updated your settings, the message
will not be re-displayed.
The following message is NOT displayed on the LCD after you try to update the settings using
the SD card: Settings updated successfully, press any key
The settings were not updated. Contact your clinician/service provider immediately.
Problem/Possible cause Solution