ResMed AutoSet CS Sleep Apnea Machine User Manual


 
English
5
Device does not stop when you remove your mask
SmartStart/Stop is disabled.
Note: SmartStart/Stop is not available if
Leak Alert is enabled.
Enable SmartStart/Stop.
SmartStart/Stop is enabled but the device does not stop automatically when you remove your
mask
Incompatible mask system being used. Only use equipment recommended by ResMed.
The patient is using a nasal pillows
mask or pediatric mask with a set
pressure less than 7 cm H
2
O.
Disable SmartStart/Stop.
Pressure rises inappropriately
Talking, coughing or breathing in an
unusual manner.
Avoid talking with a nasal mask on, and breathe as normally
as possible.
Mask cushion is buzzing against the
skin.
Adjust the headgear.
Cushion seated incorrectly causing
excessive leak.
Adjust headgear or re-fit cushion.
Displays error message: High temperature fault, refer to user manual
Device has been left in a hot
environment.
Allow to cool before re-use. Disconnect the power cord
and then reconnect it to restart the device.
Air filter is blocked. Replace your air filter. Disconnect the power cord and then
reconnect it to restart the device.
Air tubing is blocked. Check your air tubing and remove any blockages.
Disconnect the power cord and then reconnect it to restart
the device.
Humidifier setting is too high, resulting
in accumulation of water in the air
tubing.
Turn the humidifier setting down and empty the water from
the air tubing.
Problem/Possible cause Solution