RayTek MI Miniature Infrared Sensor Thermometer User Manual


 
ProgrammingGuide
74 MI
XF factorydefaultvalueswillbeset
11.6.5LockMode
Theaccesstotheunitispossibleviaserialinterface(software)and
viathedirectus erinput(modebuttons,LCDdisplay).Itispossible
tolockthebuttons.Thisallowsaccesstheunitonlyviasoftware.
J=L directuserinputvia
modebuttonsdenied
Alternativelytheunitcanbeunlockedbypressingthe<Mode/Up>
buttonssimultaneouslyforthreeseconds.
11.6.6ModeSettingfortheDigitalInputFTC3
ThedigitalinputFTC3(seesection
5.4.4TriggerandHoldFunction
onpage28)canbeusedasfollows:
XN=T FTC3astrigger
XN=H FTC3withholdfunction
11.6.7ChangingtheSensingHeadCalibrationData
Ifitisnecessarytoexchangethesensinghead,youmustsetthe
calibrationdataforthenewsensinghead:
XZ=01234567FFFFFFFF accordingtothe
headcalibrationdata
ForMID/MICmodelsthelasteightnumbersarenotusedandmust
besetto„F“likeshownintheexampleabove.For MIH modelsand
specials(likeMTBorG5)allsixteennumbersareused.
11.6.8AmbientBackgroundTemperatureCompensation
Incaseofcompensatingtheambient
backgroundtemperature,the
followingmodesareavailable:
AC=0 nocompensation
AC=1 compensationwithaconstanttemperaturevalueset
withcommandA.