Maintenance
MI 63
9.3SensingHeadExchange
Sensing heads and electronic boxes can only be
interchangedinaccordancetothefollowing table!
MID02 MIC02 MID10 MIC10 MIH10 MID20 MIC20 MIH20
MID02 x x x x
MIC02 x x x x
MID10 x x x x
MIC10 x x x x
MIH10 x
MID20 x
MIC20 x
MIH20 x
The head exchange requires to type in the new sensing head
calibrationdataprintedonthecableasfollows:
1.Toexchangethesensinghead,disconnectthepowerofthe
unit.
2.Connectthewiresforthenewsensingheadaccordingtothe
colordescriptionontheprintedcircuitboard.
3.
SwitchthepowerfortheunittoON.
4.Presssimultaneouslythe<Mode/Down/Up>buttons.
5.Fourcharactersappearinthedisplay(formervalues).Typein
thenewdesig nator(A)usingthe<Down/Up>buttons.Press
the<Mode>button.
6.Thesecondblock offour charactersappears inthe display
(former values).
Type in the new designator (B) using the