RayTek MI Miniature Infrared Sensor Thermometer User Manual


 
Maintenance
MI 63
9.3SensingHeadExchange
Sensing heads and electronic boxes can only be
interchangedinaccordancetothefollowing table!
MID02 MIC02 MID10 MIC10 MIH10 MID20 MIC20 MIH20
MID02 x x x x
MIC02 x x x x
MID10 x x x x
MIC10 x x x x
MIH10 x
MID20 x
MIC20 x
MIH20 x
The head exchange requires to type in the new sensing head
calibrationdataprintedonthecableasfollows:
1.Toexchangethesensinghead,disconnectthepowerofthe
unit.
2.Connectthewiresforthenewsensingheadaccordingtothe
colordescriptionontheprintedcircuitboard.
3.
SwitchthepowerfortheunittoON.
4.Presssimultaneouslythe<Mode/Down/Up>buttons.
5.Fourcharactersappearinthedisplay(formervalues).Typein
thenewdesig nator(A)usingthe<Down/Up>buttons.Press
the<Mode>button.
6.Thesecondblock offour charactersappears inthe display
(former values).
Type in the new designator (B) using the