RayTek MI Miniature Infrared Sensor Thermometer User Manual


 
ProgrammingGuide
MI 73
11.6DeviceControl
11.6.1OutputfortheTargetTemperature
Thesignaloutputcanbesetto420mA,020mAormV.Ifcurrent
outputisactivated,theoutputcanprovideapredefinedcurrent:
XO=4 outputmodeto420mA
O=13.57 outputofaconstantcurrentat13.57
mA
O=60 switchesbacktothetemperaturecontrolledoutput
11.6.2AnalogOutput,Scaling
Accordingtothetemperaturerangeofthemodel,itispossibletoset
amaximumvoltage/currentvalueaccordingtoatemperaturevalue
(e.g.,themaximumcurrent20mAshallrepresent200°C/ 392°F).The
samesettingis
possiblefortheminimumvalue.
H=500 themaximumcurrent/voltagevalueissetto500°C
L=0 theminimumcurrent/voltagevalueissetto0°C
Remark:Youcannotset thisvalueforthermocoupleoutput.The
minimumspanbetweenthemaximum/minimumsettingsis20K.
11.6.3AlarmOutput
Thesecondoutp utchannel
canbesetindifferentmodes,seesection
5.3.2HeadAmbientTemp./AlarmOutputonpage20.
Internalsensingheadtemperature
Alarmoutput
K=7 internalsensingheadtemperature
K=4 alarmoutput for objecttemperature,0 V in caseof no
alarm
XS=125.3thresholdsettingto125.3°C(ifU=Cisset)
11.6.4Factorydefaultvalues
Itispossibletoresettheunittothedefault
values.