RayTek MI Miniature Infrared Sensor Thermometer User Manual


 
Installation
16 MI
5.2.2CableforPowerSupplyandOutputs
Youneedtoconnectthepowersupply(12to26VDC)andthesignal
outputwires.Useonlycablewithoutsidediameterfrom4to6mm
(0.16to0.24in),AWG24.
Thecablemustincludeshieldedwires.Itshouldnotbe
used
asastrainrelief!
1. Cutabout40mm(1.5in)ofthecablesheath(7)fromtheend
ofthecable.Caution:Donotcutintotheshield!
2. Cuttheshield(5)soabout5mm(0.2in)remainsexposed
from underthe cable sheath (7). Separatethe
shield and
spreadthestrandsout.Shortentheinsideinsulationuntil
youcanseparatethewires(6).
3. Strip3mm(0.15in)ofinsulationfromthewires(6).
Figure5:CablePreparation
4. OpentheelectronicboxbyremovingthefourPhillipshead
screws andpullingoff the lid.Unscrew the cap (1),and
removetheplasticcompressionfitting(2),therubberwasher
(3),whichisinsidethefitting,andthetwometalwashers(4).