RayTek MI Miniature Infrared Sensor Thermometer User Manual


 
Operation
MI 37
Display Mode Range
C
Target Temperature* (effected
by signal processing)
not adjustable
A Head Ambient Temperature not adjustable
T
Target Temperature (not
effected by signal processing)
not adjustable
Output Mode mV mV output (default)
TCK thermocouple type K output
TCJ thermocouple type J output
4 - 20 4 - 20 mA current loop
0 - 20 0 - 20 mA current loop
E Emissivity 0.100 ... 1.000 (default: 0.950)
T Transmission 0.100 ... 1.000 (default: 1.000)
A Signal processing: Average
**
0.100 ... 999.0
P Signal processing: Peak Hold
**
0.100 ... 998.9 999 = infinite (P )
V Signal processing: Valley Hold
**
0.100 ... 998.9 999 = infinite (V )
L Low end of range L = -40 ... 600**** (default: 0)
H High end of range H = -40 ... 600**** (default: 500)
U Temperature Unit °C or °F (default: °C)
M Multidrop Address
***
1 – 32, --- for address 0 (single unit)
* appears automatically after 10 s without any action
** not simultaneously
*** only for units with RS485 interface
**** temperatures according to LT head
Table5:AvailableModes